Velleman DVM2000 Instrukcja obsługi

Kategoria
Multimetry
Typ
Instrukcja obsługi
DVM2000
HIGH-ACCURACY TRUE RMS MULTIMETER - 50000 COUNTS
TRUE RMS-MULTIMETER - ZEER NAUWKEURIG - 50000 COUNTS
MULTIMÈTRE TRUE RMS HAUTE PRÉCISION - 50000 POINTS
MULTÍMETRO TRUE RMS DE ALTA PRECISIÓN - 50000 CUENTAS
SEHR GENAUES TRUE RMS MULTIMETER - 50000 ZÄHLUNGEN
WYSOKIEJ JAKOŚCI MIERNIK Z TRUE RMS - 50000 COUNTS
USER MANUAL
GEBRUIKERSHANDLEIDING
NOTICE D’EMPLOI
MANUAL DEL USUARIO
BEDIENUNGSANLEITUNG
INSTRUKCJA OBSŁUGI
DVM2000 VELLEMAN
3
DVM2000 – HIGH-ACCURACY TRUE RMS MULTIMETER - 50000 COUNTS
1. Introduction
To all residents of the European Union
Important environmental information about this product
This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment.
Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling.
This device should be returned to your distributor or to a local recycling service.
Respect the local environmental rules.
If in doubt, contact your local waste disposal authorities.
Thank you for choosing Velleman! Please read the manual thoroughly before bringing this device into service. If the device was damaged in transit, don't
install or use it and contact your dealer.
2. Safety Instructions
This multimeter has been designed according to IEC-1010-1 concerning electronic measuring instruments with an overvoltage category CAT III 1000V
and pollution 2. This multimeter complies with 89/336/EEC, 73/23/EEC and 93/68/EEC.
Damage caused by disregard of certain guidelines in this manual is not covered by the warranty and the dealer will not accept responsibility for any
ensuing defects or problems.
Check the meter before every use. Never use a damaged or deformed multimeter.
Do not use the multimeter in an environment with combustible gases, steam or dust.
Always be careful when handling voltages > 60VDC or > 50VAC RMS. Always keep fingers behind the guards. Keep your body insulated when
measuring.
Never measure voltages exceeding 1000VAC or 1000VDC to ensure your safety.
Replace the batteries when the symbol appears. Disconnect the measuring probes from any circuit before opening the multimeter.
Repair and calibration must be carried out by experienced professionals only.
DVM2000 VELLEMAN
4
3. Description
a. Multimeter
1. V//Hz terminal
2. COM terminal
3. µA/mA terminal
4. A terminal
5. rotary selection switch
6. power switch
7. RANGE: range selection key
8. SELECT: measuring mode selection key
9. WAKE: meter reactivation key
10. MAX/MIN: maximum/minimum value display selection key
11. REL: relative measuring key
12. HOLD: data-hold key
13. LIGHT: backlight key
14. ~Hz: linear frequency measurement
15. LCD display
b. LCD Display
DVM2000 VELLEMAN
5
1
µnF
Measurement units.
mµA
dBmV
MKHz
2 Л% Duty cycle measurement.
3 AUTO Automatic range selection.
4 MANUAL Manual range selection.
5 51000 Range indication.
6 - Maximum/minimum value display.
7 Relative measurement mode.
8 Continuity measurement.
9 AC AC measurement mode.
10 Negative value.
11 DC DC measurement mode.
12 Data-hold function.
13 Please wait… Waiting mode.
14 Low battery.
15 Analogue bar indication.
16 Multimeter is linked to a PC via USB.
17 Displayed value.
4. Operation
Backlight
Press LIGHT to switch the backlight on or off. The backlight will automatically go out after 30 seconds. Remember that using the backlight will drain
the batteries faster.
DVM2000 VELLEMAN
6
Data Hold
Press HOLD to freeze the displayed value. Press again to resume measurement.
Automatic Shutoff
The meter shuts off after 15 minutes of inactivity by default. When shut off, press WAKE or turn the rotary selection switch to reactivate the
multimeter. To deactivate the automatic shutoff function, press the power switch and WAKE simultaneously. Remember to shut off the multimeter
using the power switch after use as the multimeter will still drain batteries when in standby.
Relative Value Measurement
This function applies for all measurement except for frequency, duty cycle and diode measurements. Press REL to determine the relative value.
The next displayed value represents the present measured value minus the recorded relative value. Press again to exit the function.
Maximum/Minimum Value Measurement
This function applies for all measurement except for frequency and diode measurements. Press MAX/MIN to determine the maximum, the minimum
or both values respectively. Hold MAX/MIN pressed for 2 seconds to exit the function.
Connecting the Multimeter to a Computer
Connect your DVM2000 to a computer using the included optical USB cable. Hold HOLD pressed for 2 seconds to send the data to the computer.
You will now be able to edit the recorded data on your computer using the appropriate software. Hold HOLD pressed for 2 seconds to stop sending
the data.
ACV/dBm Measurements
Voltage ranges: 0.5VAC~1000VAC
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
Do not measure voltages exceeding 1000V.
DVM2000 VELLEMAN
7
o Select the ACV or dBm measurement mode by pressing SELECT.
o Connect the measuring probes to the circuit.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
DCV/DCV + ACV Measurements
Voltage ranges: 0.5VAC~1000VAC
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
o Select the DCV or DCV + ACV measurement mode by pressing SELECT.
o Connect the measuring probes to the circuit.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
DC mV/AC mV/DC mV + AC mV Measurements
Voltage ranges: 1µV~500mV
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
o Select the DC mV, AC mV or DC mV + AC mV measurement mode by pressing SELECT.
o Connect the measuring probes to the circuit.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
Do not measure voltages exceeding 1000V.
DVM2000 VELLEMAN
8
Logic Frequency/Duty Cycle Measurements
Frequency ranges: 5Kz~2MHz (Vp 2.5V~5V), duty cycle ranges: 5~95%
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
o Select the Hz or % measurement mode by pressing SELECT.
o Connect the measuring probes to the circuit.
o Read the measured value. A value too low will be displayed as 0.
Diode Measurements
Diode ranges: 0V~2.5V
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
o Connect the red measuring probe to the positive pole and the black probe to the negative pole of the diode. The
LCD will display the forward voltage drop.
o Connect the black measuring probe to the positive pole and the red probe to the negative pole of the diode. The
backward resistance of the diode is normal when OL is displayed.
o Remember to turn off the circuit before measuring diodes integrated in that circuit.
Resistance/Continuity Measurements
Resistance ranges: 0.01~50M
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
DVM2000 VELLEMAN
9
o Select the or measurement mode by pressing SELECT.
o Connect the measuring probes to the resistor and read the measured resistance value.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
o Simply connect the measuring probes to the circuit for the continuity test.
o Remember to turn off the circuit before measuring resistance or continuity.
Capacitance Measurements
Capacitance ranges: 10pF~5000µF
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the V//Hz and the COM terminals respectively.
o Discharge the capacitor before measurements.
o Connect the red measuring probe to the positive pole and the black measuring probe to the negative pole of a
heteropolar capacitor.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
DC µA/AC µA/DC µA + AC µA Measurements
Current ranges: 0.01µA~5000µA
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the mA/µA and the COM terminals respectively.
o Select the DC µA, AC µA or DC µA + AC µA measurement mode by pressing SELECT.
o Turn off the power to the circuit to be measured. Connect the measuring probes in series to the circuit and turn on
the power to the circuit.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
DVM2000 VELLEMAN
10
DC mA/AC mA/DC mA + AC mA Measurements
Current ranges: 1µA~500mA
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red and the black measuring probes in the mA/µA and the COM terminals respectively.
o Select the DC mA, AC mA or DC mA + AC mA measurement mode by pressing SELECT.
o Turn off the power to the circuit to be measured. Connect the measuring probes in series to the circuit and turn on
the power to the circuit.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
DC A/AC A/DC A + AC A Measurements
Current ranges: 0.1mA~10A
o Switch on your multimeter and set to rotary selector switch to .
o Insert the red measuring probe in the A terminal and the black measuring probe in the COM terminal.
o Select the DC A, AC A or DC A + AC A measurement mode by pressing SELECT.
o Turn off the power to the circuit to be measured. Connect the measuring probes in series to the circuit and turn on
the power to the circuit.
o Read the measured value. Select a higher range by pressing RANGE when ol is displayed.
Linear Frequency Measurements
Frequency ranges: 5Hz~200kHz
o Press ~Hz to measure the linear frequency when performing voltage or current measurements.
o Select the range by pressing RANGE.
DVM2000 VELLEMAN
11
Range Sensitivity (sine wave)
500mV 100mV
5V 0.5V
50V 4V
500V 40V
1000V 400V
5000µA 1mA
500mA 100mA
o Press ~Hz to exit the function.
5. Cleaning and Maintenance
Battery Replacement
The symbol appears on the LCD as soon as the total voltage of the batteries drops below 6.8V. Before replacement, disconnect the measuring
probes from the circuit and turn off the power to the meter. Remove the back cover using an appropriate Philips-head screwdriver. Replace the batteries
with batteries of the same specifications respecting the polarity. Close the back cover before use.
Fuse Replacement
Before replacement, disconnect the measuring probes from the circuit and turn off the power to the meter. Remove the back cover using an appropriate
Philips-head screwdriver. Replace the fuse with a fuse of the same specifications. Make sure to insert the fuse correctly. Close the back cover before
use.
Meter Calibration
Calibration must only be executed by qualified professionals. Please contact your dealer.
Other
Wipe the device regularly with a dry, lint-free cloth. Do not use alcohol or solvents. Keep the multimeter away from rain and moisture.
Never use a damaged or deformed multimeter. Please contact your dealer for repairs.
DVM2000 VELLEMAN
12
6. Technical Specifications
Accuracy is specified for one year after calibration at operating temperatures of 18°C ~ 28°C with a relative humidity at 0% ~ 80%. Accuracy
specifications take the form of ± (% of rdg + number of least significant digits).
Environmental Conditions 1000V CAT III
Pollution Degree 2
Altitude < 2000m
Operating Temperature 0°C~30°C or 32°F~86°F (< 80% RH, < 10°C)
Storage Temperature -20°C~60°C or -4°F~140°F (< 70% RH, batteries removed)
Fuse µA and mA, F630mA / 500V
A, F12.5A / 500V
Display 99999
Over Range Indication yes (“OL”)
Low Battery Indication yes ( )
Data Hold yes
Backlight yes
Power Supply 6 x AAA batteries
Dimensions 205 x 97 x 45mm
Weight ± 560g (with batteries)
Accessories 2 x test leads, 1 x optical USB cable, PC-link software, 6 x AAA batteries, manual
AC Voltage/DC + AC Voltage
Range Resolution Accuracy
40Hz~1kHz 1kHz~10kHz 10kHz~20kHz
50mV 0.001mV
± (0.5% + 40) ± (1% + 40) ± (2.5% + 40)
500mV 0.0mV
5V 0.1mV
DVM2000 VELLEMAN
13
50V 1mV
± (0.5% + 40) ± (1% + 40) ± (2.5% + 40)
500VV 10mV unspecified
1000V 0.1V unspecified unspecified
Above accuracies guaranteed within 10%~100% of the full range.
DC Voltage
Range Resolution Accuracy
50mV 0.001mV ± (0.03% + 10)
500mV 0.01mV
± (0.03% + 6)
5V 0.1mV
50V 1mV
500V 10mV
1000V 0.1V
Above accuracies guaranteed within the full range.
AC Current/DC + AC Current
Range Resolution Accuracy Voltage Drop
40Hz~1kHz 1kHz~10kHz 10kHz~20kHz
500µA 0.01µA ± (0.75% + 20) ± (1% + 20) ± (2% + 20) 102µV/µA
5000µA 0.1µA ± (0.75% + 10) ± (1% + 10) ± (2% + 10)
50mA 1µA ± (0.75% + 20) ± (1% + 20) ± (2% + 20) 1.5mA/mA
500mA 10µA ± (0.75% + 10) ± (1% + 10) ± (2% + 10)
5A 0.1mA ± (0.75% + 20) ± (1.5% + 20) ± (5% + 20) 30mV/A
10A 1mA ± (1.0% + 10) ± (1.5% + 10) unspecified
Above accuracies guaranteed within 10%~100% of the full range.
DVM2000 VELLEMAN
14
DC Current
Range Resolution Accuracy Voltage Drop
500µA 0.01µA ± (0.15% + 15) 102µV/µA
5000µA 0.1µA
± (0.15% + 10)
50mA 1µA 1.5mA/mA
500mA 10µA
5A 0.1mA
± (0.5% + 10) 30mV/A
10A 1mA
Above accuracies guaranteed within the full range.
Resistance
Range Resolution Accuracy
500 0.01 ± (0.1% + 10)
5k 0.1
± (0.1% + 5)
50k 1
500k 10
5M 100 ± (0.1% + 10)
50M 1k ± (0.5% + 10)
Above accuracies guaranteed within the full range.
Capacitance
Range Resolution Accuracy
50nF 0.01nF
± (1% + 5)
500nF 0.1nF
5µF 1nF
50µF 10nF
500µF 0.1µF
± (2% + 5)
5000µF 1µF
Above accuracies for film capacitors or better guaranteed within the full range.
DVM2000 VELLEMAN
15
Diode
Range Resolution Accuracy
2.5V 0.1mV ± (1% + 5)
Test current 0.7mA
Logic Frequency
Frequency Range Sensitivity Accuracy
5Hz~2MHz Vp 2~5V square wave ± (0.006% + 4)
Linear Frequency
Frequency Range Voltage/Current Range Sensitivity Accuracy
5Hz~200kHz
(sine wave)
500mV 100mV
± (0.006% + 4)
5V 0.5V
50V 4V
500V 40V
1000V 400V
5000µV 1mV
500mV 100mV
Accuracy lowers for low voltage or low frequency.
Duty Cycle
Frequency Range Duty Cycle Range Resolution Accuracy
5Hz~500kHz 5%~95% 0.01% ± (2% + 5-
Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage or injury resulted from
(incorrect) use of this device.
For more info concerning this product, please visit our website www.velleman.eu.
The information in this manual is subject to change without prior notice.
DVM2000 VELLEMAN
16
DVM2000 – TRUE RMS-MULTIMETER - ZEER NAUWKEURIG - 50000 COUNTS
1. Inleiding
Aan alle ingezetenen van de Europese Unie
Belangrijke milieu-informatie betreffende dit product
Dit symbool op het toestel of de verpakking geeft aan dat, als het na zijn levenscyclus wordt weggeworpen, dit toestel schade kan toebrengen
aan het milieu.
Gooi dit toestel (en eventuele batterijen) niet bij het gewone huishoudelijke afval; het moet bij een gespecialiseerd bedrijf terechtkomen voor
recyclage.
U moet dit toestel naar uw verdeler of naar een lokaal recyclagepunt brengen.
Respecteer de plaatselijke milieuwetgeving.
Hebt u vragen, contacteer dan de plaatselijke autoriteiten inzake verwijdering.
Dank u voor uw aankoop! Lees deze handleiding grondig voor u het toestel in gebruik neemt. Werd het toestel beschadigd tijdens het transport,
installeer het dan niet en raadpleeg uw dealer.
2. Veiligheidsinstructies
Deze multimeter werd ontworpen volgens IEC-1010 aangaande elektrische metingen met een overspanning CAT III 1000V en vervuiling 2. Deze
multimeter voldoet aan 89/336/EEC, 73/23/EEC en 93/68/EEC.
De garantie geldt niet voor schade door het negeren van bepaalde richtlijnen in deze handleiding en uw dealer zal de verantwoordelijkheid afwijzen
voor defecten of problemen die hier rechtstreeks verband mee houden.
Controleer de meter voor elk gebruik. Gebruik geen beschadigde of vervormde multimeter.
Vermijd gebruik in een omgeving met brandbare gassen, stoom of stof.
Wees voorzichtig wanneer u werkt met spanningen > 60VDC of > 50VAC rms. Houd uw vingers altijd achter de beschermingen.
Voor uw eigen veiligheid, meet nooit spanningen boven 1000VAC of 1000VDC.
Vervang de batterijen wanneer verschijnt. Ontkoppel de meetsnoeren van het circuit alvorens de multimeter te openen.
Vertrouw reparaties en ijking enkel toe aan professionelen.
DVM2000 VELLEMAN
17
3. Omschrijving
a. Multimeter
1. V//Hz terminal
2. COM terminal
3. µA/mA terminal
4. A terminal
5. draaischakelaar
6. voedingsschakelaar
7. RANGE: keuzetoets bereik
8. SELECT: keuzetoets meetmodus
9. WAKE: reactiveringtoets
10. MAX/MIN: keuzetoets maximum-/minimumwaarde
11. REL: keuzetoets relatieve meting
12. HOLD: data-holdtoets
13. LIGHT: achtergrondverlichting
14. ~Hz: meting lineaire frequentie
15. lcd-scherm
b. Lcd-scherm
DVM2000 VELLEMAN
18
1
µnF
Meeteenheden.
mµA
dBmV
MKHz
2 Л% Meting cyclische verhouding.
3 AUTO Automatische bereikinstelling.
4 MANUAL Manuele bereikinstelling.
5 51000 Bereikaanduiding.
6 - Weergave maximum-/minimumwaarde.
7 Relatieve meetmodus.
8 Doorverbindingmeting.
9 AC AC-meting.
10 Negatieve waarde.
11 DC DC-meting.
12 Data-holdfunctie.
13 Please wait… Wachten.
14 Zwakke batterij.
15 Analoge balk.
16 De multimeter is aan een pc aangesloten via usb.
17 Weergegeven waarde.
4. Gebruik
Achtergrondverlichting
Druk op LIGHT om de achtergrondverlichting in of uit te schakelen. De achtergrondverlichting schakelt na 30 seconden automatisch uit. Merk dat de
batterijen sneller uitputten wanneer u de achtergrondverlichting vaak gebruikt.
DVM2000 VELLEMAN
19
Data hold
Druk op HOLD om de aflezing op het scherm vast te spijkeren. Druk opnieuw om de meting te hervatten.
Automatische uitschakeling
De meter wordt na 15 minuten inactiviteit automatisch uitgeschakeld. Druk op WAKE of draai aan de draaischakelaar om de multimeter opnieuw in
te schakelen. Schakel de automatische uitschakeling uit door de voedingsschakelaar en WAKE gelijktijdig in te drukken. Schakel de meter na
gebruik uit via de voedingsschakelaar aangezien een multimeter in stand-by toch verbruikt.
Relatieve metingen
Deze functie kan gebruikt worden voor alle meetfuncties uitgenomen voor frequentiemeting, diodemeting en meting van cyclische verhouding. Druk
op REL om de relatieve waarde vast te leggen. De weergegeven waarde stelt nu de huidige gemeten waarde min de vastgelegde relatieve waarde.
Druk opnieuw om de functie te verlaten.
Maximum-/minimumwaarde
Deze functie kan gebruikt worden voor alle meetfuncties uitgenomen voor frequentie- en diodemeting. Druk op MAX/MIN om de maximum-, de
minimum- of beide waarden te bepalen. Houd MAX/MIN gedurende 2 seconden ingedrukt om de functie te verlaten.
De multimeter aan een computer koppelen
Koppel uw DVM2000 aan een pc met behulp van de meegeleverde optische usb-kabel. Houd HOLD gedurende 2 seconden ingedrukt om de
gegevens naar de computer door te zenden. Met de juiste software kunt u nu de gegevens editen. Houd HOLD gedurende 2 seconden ingedrukt om
de gegevensoverdracht naar de pc te stoppen.
Meten van ACV/dBm
Spanningsbereik: 0.5VAC~1000VAC
o Schakel de multimeter in en plaats de draaischakelaar op .
o Verbind het rode en het zwarte meetsnoer met de V//Hz respectievelijk de COM terminal.
Meet geen spanningen groter dan 1000V.
DVM2000 VELLEMAN
20
o Selecteer de ACV of dBm meetmodus door op SELECT te drukken.
o Verbind de meetsnoeren met het circuit.
o Lees de weergegeven waarde. Selecteer een hoger bereik met RANGE wanneer ol verschijnt.
Meten van DCV/DCV + ACV
Spanningsbereik: 0.5VAC~1000VAC
o Schakel de multimeter in en plaats de draaischakelaar op .
o Verbind het rode en het zwarte meetsnoer met de V//Hz respectievelijk de COM terminal.
o Selecteer de DCV of DCV + ACV meetmodus door op SELECT te drukken.
o Verbind de meetsnoeren met het circuit.
o Lees de weergegeven waarde. Selecteer een hoger bereik met RANGE wanneer ol verschijnt.
Meten van DC mV/AC mV/DC mV + AC mV
Spanningsbereik: 1µV~500mV
o Schakel de multimeter in en plaats de draaischakelaar op .
o Verbind het rode en het zwarte meetsnoer met de V//Hz respectievelijk de COM terminal.
o Selecteer de DC mV, AC mV of DC mV + AC mV meetmodus door op SELECT te drukken.
o Verbind de meetsnoeren met het circuit.
o Lees de weergegeven waarde. Selecteer een hoger bereik met RANGE wanneer ol verschijnt.
Meet geen spanningen groter dan 1000V.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80

Velleman DVM2000 Instrukcja obsługi

Kategoria
Multimetry
Typ
Instrukcja obsługi