EMC Declaration
IEC 60601 -1 -2:2014 Tesla Smart Blood Pressure Monitor: iden-
tification, marking and documents for Class B product
Instructions for use
The Tesla Smart Blood Pressure Monitor is suitable for home
healthcare environments and so on.
Warning: Don‘t hear active HF surgical equipment and the RF
shielded room of a system for magnetic resonance imaging,
where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with
other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are
operating normally,
Warning: Use of accessories, transducers and cables other
than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and
result in improper operation.
Warning: Portable RF Communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of
the blood pressure monitor, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this
equipment could result.
If any: a list of all cables and maximum lengths of cables (if appli-
cable), transducers and other accessories that are replaceable
by the responsible organization and that are likely to affect Tesla
Smart Blood Pressure Monitor with the requirements of Clause
7 (EMISSIONS) and Clause 8 (IMMUNITY). Accessories may
be specified generically (e.g., Shielded cable, load impedance)
or specifically (e.g., by manufacturer and equipment or type
reference].
If any: the performance of Tesla Smart Blood Pressure Monitor
that was determined to be essential performance and a descrip-